Solid-State Nuclear Track Detectors (SSNTD)

A solid-state nuclear track detector (SSNTD) is a solid insulating material — a plastic, a glass or a mineral crystal — that records individual charged particles as permanent damage trails. Chemical etching enlarges each trail into a pit you can count and measure under a microscope. You will also see them called nuclear track detectors (NTDs) or etched track detectors.

The most widely used SSNTD is CR-39, a plastic also known as PADC. This guide explains how SSNTDs work, compares the common materials, and shows which one fits which measurement. BlankSlate Innovation makes CR-39 SSNTDs in-house in Lubbock, Texas.

How it works

How Does an SSNTD Record a Particle?

Every SSNTD works the same way, whatever it is made of. The detector does nothing during the exposure except accumulate damage. All the measurement happens afterwards.

01

Latent Track

A charged particle passes through the solid and breaks chemical bonds along a trail a few nanometers wide. The trail is permanent but invisible, and the detector needs no power to record it.

02

Chemical Etch

An etchant dissolves the damaged material faster than the undamaged bulk. Each trail opens into a cone-shaped pit microns across. Plastics are etched in hot NaOH or KOH; mica and glass need hydrofluoric acid.

03

Readout

Pits are counted and measured with an optical microscope, an automated reader or a scanning electron microscope. The number of pits gives particle fluence. Their size and shape carry information about energy loss, particle type and direction.

Only particles that deposit enough energy per unit path length (linear energy transfer, LET) leave an etchable trail. That threshold is what separates one SSNTD material from another, and it is why no track detector responds to gamma rays, X-rays or beta particles. For a plain-language walk-through, see What Is CR-39?

Side by side

SSNTD Materials Compared

Typical values from the track-detector literature. Exact response depends on the product, the etch and the readout, so treat this table as a starting point and calibrate on your own setup.

Material What it is Records Does not record Typical etch Common uses
CR-39 / PADC Clear thermoset plastic sheet, poly(allyl diglycol carbonate) Protons, alpha particles, heavy ions, fission fragments; neutrons through recoils or a converter coating Gamma rays, X-rays, beta particles Hot NaOH or KOH. BSI's reference is 6.25 M NaOH at 80 °C Radon, neutron dosimetry, fusion and laser-plasma diagnostics, cosmic rays, teaching labs
LR-115 (cellulose nitrate) Thin red-dyed cellulose nitrate layer, about 12 µm, on a polyester base Alpha particles within a limited energy window, roughly 1.5 to 4 MeV Alpha particles outside that window, most protons, gamma rays, X-rays, beta particles About 2.5 N NaOH at 60 °C Radon and thoron surveys, alpha autoradiography
Polycarbonate (Makrofol, Lexan) Bisphenol-A polycarbonate film or sheet Fission fragments and heavy ions; alpha particles only with electrochemical etching Protons, gamma rays, X-rays, beta particles KOH in ethanol and water, often with an applied field Fission-track neutron dosimetry, heavy-ion experiments
Muscovite mica Natural mineral crystal, cleaved into thin sheets Fission fragments and very heavy ions only Alpha particles, protons and other light ions Hydrofluoric acid Fission-track dating, uranium mapping, heavy-ion counting in a strong alpha background
Glass (soda-lime, phosphate) Inorganic glass plate or slide Fission fragments and heavy ions Alpha particles, protons and other light ions Hydrofluoric acid Fission-fragment counting, high-dose and high-temperature environments

BSI manufactures CR-39 (PADC). We do not sell LR-115, polycarbonate, mica or glass detectors; they are here so the comparison is complete. LR-115, Makrofol and Lexan are product names of their respective owners.

The default choice

Why Is CR-39 the Most Used SSNTD?

Lowest Threshold

CR-39 is the most sensitive common SSNTD and the one that records protons. BSI's working threshold is about 0.8 MeV/(mg/cm2) in dE/dx, roughly 50–100 keV for most ions.

One Detector, Many Particles

Protons, alphas, heavy ions and fission fragments all register. Track diameter relates to LET, so one chip can yield species and energy information as well as a count.

Clear and Uniform

The plastic is optically transparent and cast in uniform sheets, which makes microscopy straightforward and lets detectors be cut to any holder or stacked.

Neutrons, With a Coating

Fast neutrons register through recoil protons. A boron nitride converter coating adds thermal and epithermal response through the 10B(n,α)7Li reaction.

Where an SSNTD is the wrong tool. Track detectors are time-integrated: you etch and read after the exposure, so they cannot tell you when a particle arrived. They do not see gamma rays, X-rays or beta particles. Mixed fields are hard to unravel without controls, because a 1 MeV alpha particle can look much like a 300 keV proton. If you need timing or a live count rate, pair the SSNTD with an electronic detector.

BSI's CR-39 work is peer-reviewed: see our papers on SEM large-area mapping with AI track analysis and boron nitride neutron coatings, or the full list on the research page.

Questions people ask

SSNTD FAQ

What does SSNTD stand for?

SSNTD stands for solid-state nuclear track detector. It is a solid insulating material, usually a plastic, glass or mineral crystal, that records the passage of individual charged particles as permanent damage trails. Chemical etching enlarges each trail into a pit that can be counted and measured under a microscope. The same devices are also called nuclear track detectors (NTDs) or etched track detectors.

Is CR-39 an SSNTD?

Yes. CR-39 is the most widely used solid-state nuclear track detector. It is a thermoset plastic, poly(allyl diglycol carbonate) or PADC, so it is also called a plastic nuclear track detector (PNTD). CR-39 is sensitive enough to record protons as well as alpha particles, heavy ions and fission fragments, which is why it is the default choice for most new work.

What is the difference between SSNTD, PNTD, PADC and CR-39?

SSNTD is the whole category: any solid that records particle tracks, including plastics, glasses and minerals. PNTD means a plastic SSNTD. PADC, poly(allyl diglycol carbonate), is the specific polymer most PNTDs are made of, and CR-39 is the original trade name for that polymer. In practice, a paper that says CR-39 detector, PADC detector, PNTD or SSNTD is usually describing the same material.

Which SSNTD material should I use?

For most work, CR-39 (PADC), because it is the most sensitive of the common materials and records protons, alpha particles and heavier ions. LR-115 cellulose nitrate film is used mainly for radon and alpha counting. Polycarbonate, mica and glass record only heavier, more densely ionizing particles such as fission fragments, which is useful when a background of lighter particles must be ignored.

Can an SSNTD detect gamma rays, X-rays or beta particles?

No. Track detectors respond only to densely ionizing particles. Gamma rays, X-rays and beta particles deposit too little energy per unit path length to leave an etchable track. That is often an advantage: an SSNTD can count alpha particles, protons or neutron-induced recoils in a strong photon background that would swamp an electronic detector.

Where can I buy SSNTDs?

BlankSlate Innovation (BSI) manufactures CR-39 (PADC) SSNTDs in-house in Lubbock, Texas, and ships worldwide. Standard detectors cost $2 per cm² and boron-coated, neutron-sensitive detectors cost $5 per cm², with a $100 minimum order. Standard thicknesses are approximately 0.9 mm and 3 mm. Orders ship in as little as 3 business days; large orders can take up to 4 weeks.

Available here

US-Made CR-39 SSNTDs

BSI casts, cuts and polishes CR-39 (PADC) in Lubbock, Texas. Standard detectors are $2/cm2 and boron-coated detectors are $5/cm2, in ~0.9 mm and 3 mm, with a $100 minimum order. Orders ship worldwide in as little as 3 business days; large orders can take up to 4 weeks.