Noble Gas Laboratory Services

Noble Gas Isotope Analysis for Natural Gas

BlankSlate Innovation operates a noble gas isotope analysis laboratory that accepts natural gas samples from operators, land teams, and exploration companies. We measure He-3 concentration and the ³He/⁴He ratio in each submitted sample and return a quantitative report — no instrument purchase or on-site visit required.

Analytical services

What We Measure

Each natural gas sample submitted to BSI's noble gas laboratory is analyzed for the following, using isotope-resolved mass spectrometry rather than standard gas composition analysis.

He-3 Concentration

Quantitative He-3 concentration, reported down to a detection limit of 0.5 ppb.

³He/⁴He Ratio

The helium isotope ratio, used to characterize gas source and support reservoir evaluation.

Total Helium

Total helium content (He-3 + He-4) reported alongside the isotope ratio, so you get the full picture in one submission.

Results are reported to you as concentrations and ratios — a natural gas sample, a gas sampling cylinder, or copper tube submission all go through the same analytical services pipeline.

Submit a sample

Submitting a Natural Gas Sample

BSI accepts natural gas samples for noble gas isotope analysis from wellhead or separator sources. Accepted container type, required volume and pressure, and shipping and labeling instructions are confirmed with your BSI contact once you submit a request — they vary by well configuration and how the sample is collected.

Pricing and turnaround

What It Costs, and How Long It Takes

We recommend submitting 7 samples collected across multiple days rather than a single sample. Well output and gas composition can vary day to day, and a single reading can miss that variation; averaging across a week gives a more statistically representative result.

1 sample

$200

A single spot check of your gas stream.

3 samples

$500

Useful for testing multiple wells or production points in one submission.

Recommended

7 samples

$1,000

Collected across multiple days for a statistically representative result. Contact us for bulk rates above 7 samples.

2–4 weeks

Typical turnaround from the time we receive your samples to a reported result.

The scarcity problem

Why So Few Labs Offer This

Most facilities capable of He-3/He-4 isotope ratio analysis are academic or federal research labs, not commercial service providers set up to take samples from operators. Access to that capability is typically through a research collaboration, not a purchase order — which leaves most operators with no practical path to get their gas tested at all.

The difficulty is analytical, not just organizational. At mass 3, He-3 (3.0160293 amu) sits just 0.0058 amu from hydrogen deuteride, HD (3.0218324 amu). Telling the two apart requires resolving power above roughly 500 at mass 3 — well beyond what a standard quadrupole residual gas analyzer can deliver, and one reason the capability stays concentrated in a small number of specialized instruments worldwide.

Nominal m/zSpecies AMass (amu)Species BMass (amu)Δm (amu)R required
3He-33.0160293HD3.02183240.0058> 500

R required = m / Δm at 10% valley definition.

Frequently asked questions

Noble Gas Analysis FAQ

What labs measure helium-3 in natural gas?
Most facilities capable of He-3/He-4 isotope ratio analysis are academic or federal research labs, not commercial service providers set up to take samples from operators. BlankSlate Innovation is a commercial noble gas laboratory that accepts natural gas samples directly from operators and land teams and reports He-3 concentration and the ³He/⁴He ratio.
How do I sample my well for helium isotopes?
Sample container type, volume, pressure, and shipping instructions are confirmed with your BSI contact when you request an analysis, since they vary by well configuration and sample source. Submit a request below to get started.
Can a standard gas composition analysis measure He-3?
No. A standard gas chromatography (GC) composition analysis measures total helium but does not resolve the isotopes. Distinguishing He-3 from He-4, and separating He-3 from interfering HD at mass 3, requires isotope-ratio mass spectrometry.
What does He-3 isotope analysis cost?
$200 for 1 sample, $500 for 3 samples, or $1,000 for 7 samples. Contact us for bulk rates above 7 samples.
How many samples should I submit?
We recommend 7 samples collected across multiple days to average out day-to-day variation in well output and get a statistically representative result.
How long does analysis take?
Results are typically returned 2 to 4 weeks from when we receive your sample.
How much He-3 is typically present in natural gas?
It varies widely by basin and has gone unmeasured across most of the country's producing gas. Publicly reported values range from a few parts per billion to over 14 ppb in high-grade helium reservoirs; most wells have no He-3 data at all. See our natural gas He-3 page for the economics behind why this matters.